CE

The actual product may differ slightly from the image.

Alternative arc quenching and insulating gases

MultiAnalyser CA (four-in-one)

CA-3-039R-R401

Gas type
CA 
Sensors
moisture 
Sensors
O2 
Sensors
NO 
Sensors
NO2 
Features
storage of measuring results  
Features
WiFi  
Features
gas return system 

For gas quality analysis

With its innovative equipment, the MultiAnalyser CA perfectly meets the user’s requirements and is ready for operation immediately after switching on. High-quality manufacture and ergonomic design guarantee the quality standards for a compact and maintenance-friendly measuring device with high measuring accuracy. This multi-functional measuring device allows the emission-free determination of up to five measuring parameters with only one sample. Depending on the individual device configuration, values of the integrated sensors can be determined. Thanks to the easy-to-remove integrated battery, transport regulations are no longer an issue. Our field-replaceable sensors offer a great benefit as the device is ready for use immediately after replacement without any down times.

The MultiAnalyser CA allows different methods of operation for emission-free handling of the measured gas (Information on which parameters determine your device can be found in the technical data). On the one hand the internal storage of the measured gas into the device, into an external cylinder or an external discharge gas collecting bag; for continuous measurements without pumping the gas back it is recommended to collect the gas in an external discharge gas collecting bag. On the other hand pumping the gas back into an external cylinder, vessel or gas compartment up to 10 bar pe. Furthermore, the external bag can be emptied by using the MultiAnalyser CA, a DILO service cart or compressor unit.

The measuring device offers intuitive operation via a 7" touch screen. It is possible to operate and exchange data by means of mobile end devices such as smartphones, tablets or laptops via WiFi. The remaining lifetime of the electrochemical sensors is automatically displayed. Integrated into a trolley, the measuring device can be transported in a safe and comfortable way. Modern control technology in conjunction with a user-friendly interface in several languages make device operation simple and convenient.

Standard version

  • battery operation and/or external power supply
  • indication of inlet pressure in bar, psi, MPa and kPa (in pa or pe) to be selected on the touch screen
  • 6 m long connecting hose
  • 2 m long electrical connecting cable
  • USB flash drive with data file for evaluation and reading out of measuring results
Learn more

Advantages & functions

Sensor data
Sensor Measuring principle Measuring range Measuring accuracy
Moisture Electronic dew point measurement (capacitive) - 60 °C to + 20 °C ≤ ± 2 °C (at > −40 °C)
≤ ± 3 °C (at < −40 °C)
Mol percent* O2 Optical measuring principle 0 - 25 Mol-%

 

≤ ± 0,3 Mol-% (to 12%)**
≤ ± 0,5 Mol-% (to 25%)***

 

Concentration NO Electrochemical
reaction
0 – 25 ppm ≤ ± 2 % of measuring range
Concentration NO2 Electrochemical
reaction
0 - 20 ppm ≤ ± 2 % of measuring range
CO Electrochemical
reaction
0 - 500 ppm  ± 2 %

*Mol-% represents the amount of substance in a mixture and is equivalent to the ideal volume fraction. Its size is independent of pressure and temperature.
** Valid for sensor temperature of 15°C - 40°C and a pressure range of 956 mbar ±100 mbar.
*** Valid for sensor temperature of 10°C - 50°C and a pressure range of 956 mbar ±100 mbar.

Overview MultiAnalyser CA
Device Moisture O2 0-25 % NO 0-25 ppm NO2 0-20 ppm CO
R201 X X      
R401 X X X X  
R501
X X X X X

 

  • emission-free measurement
  • modular interchangeability of the sensors
  • storage of up to 500 measuring results with name, date and time
  • limit values to be set individually for each sensor
  • precise measuring results for subsequent measurements can be guaranteed by automatically purging the measuring hose prior to each measurement

Technical data

Dimensions (W x H x D) 538 x 269 x 406 mm
Weight 25 kg
Inlet pressure pe 0.2 - 35 bar
Operating temperature -10 to +50 °C
Ambient moisture max. 90 % relative moisture, non condensing
Operating voltage 85 - 264 V AC
Frequency 47 - 63 Hz
Number of max. measured values to be stored 500 
Interface USB / LAN / WiFi
Measuring time max. 7 min variable calculated by the system
Protection class IP65 (device closed) / IP20 (device opened)
Measuring principle of moisture sensor Electronic dew point measurement (capacitive)
Measuring range of moisture sensor -60 to +20 °C
Measuring accuracy of moisture sensor ≤ ± 2 °C (at > −40 °C )
Measuring accuracy of moisture sensor ≤ ± 3 °C (at < −40 °C )
Measuring principle O2 sensor Optical measuring principle
Measuring range O2 sensor 0 - 25 mol %
Measuring accuracy O2 sensor ≤ ± 0,3 mol % (to 12%)**
Measuring accuracy O2 sensor ≤ ± 0,5 mol % (to 12%)**
Measuring principle concentration NO sensor and NO2 sensor Electrochemical reaction
Measuring range concentration NO sensor and NO2 sensor 0 – 25 ppm NO
Measuring range concentration NO sensor and NO2 sensor 0 - 20 ppm NO2
Measuring accuracy concentration NO sensor and NO2 sensor ≤ ± 2 % of measuring range

Optional accessories

Compressor unit for measuring devices 3-826-R003
Learn more Add to cart
6 m long connecting hose with self-closing DILO couplings (as extension hose) 3-531-R060
Learn more Add to cart
Discharge gas collecting bag B151R96
Learn more Add to cart
Hose connection DN8 with thread (M24x1,5) CA-6-1161-R011
Learn more Add to cart
DILO coupling DN40 (M50x2) with hose connection CA-6-1161-R012
Learn more Add to cart